    How should you handle common faults in a high-low temperature test chamber?

    Source:LINPIN Time:2025-08-20 Category:Industry News

    High-low temperature test chambers are precision instruments that must be operated strictly by trained personnel following standard procedures and subjected to regular preventive maintenance. Any operational error or neglected upkeep can trigger a variety of faults, delaying test schedules. The equipment integrates refrigeration, heating, humidification, electrical control and other subsystems. Effective troubleshooting therefore requires a holistic “whole-machine mindset” to pinpoint and solve problems quickly.

    1 Slow cooling or no cooling at all
    1.1 First confirm that set-point parameters are correct. Then listen to the compressor and monitor its running current. If the compressor does not start, inspect the power supply, contactor, thermal relay, start capacitor and control-board outputs.
    1.2 If the compressor runs but the high-side pressure is too high while the low-side pressure is too low, the condenser is usually fouled or the condenser fan has failed. Clean the condenser fins or replace the fan.
    1.3 If both high- and low-side pressures are low and frost appears on the suction line, suspect refrigerant leakage or an iced-up expansion valve. Use an electronic leak detector to locate leaks, repair them, evacuate the system and recharge with the correct refrigerant. Replace the filter-drier and re-evacuate if the expansion valve is blocked by ice.
    1.4 When the condenser is iced, perform a manual defrost cycle or temporarily raise the chamber temperature to melt the ice before resuming normal operation.

    High-low temperature test chambers

    2 Heating or humidification anomalies
    2.1 Observe the actual vs. set-point values on the touchscreen. If the deviation is large and the power output reads 0, first check whether the solid-state relay (SSR) has failed short or open. Measure the SSR output terminals with a multimeter.
    2.2 If the SSR is sound but the heater or humidifier still carries no current, disconnect the heater and measure its resistance; an infinite reading indicates replacement. Clean or replace a scaled-up humidifier tube.
    2.3 Verify that the water-level switch and feed-water solenoid valve are functional to prevent dry-fire of the humidifier.

    3 Door “sticks” shut in rainy season
    High humidity can create a vacuum between the door and the silicone gasket. Wipe off moisture with a dry cloth before closing the door and apply a thin film of talcum powder or silicone grease to the gasket for both moisture-proofing and lubrication. Replace the gasket if it is cracked or aged.

    4 No response when pressing the start button
    4.1 With a multimeter confirm that the mains voltage is within ±10 % of the rated value. Check the main breaker, emergency-stop button and fuses inside the chamber.
    4.2 If ambient temperature falls below 5 °C in winter, the compressor’s low-temperature protection may activate. Turn on room air-conditioning or another heater to raise the ambient temperature above 10 °C before restarting the chamber.

    5 Sudden increase in noise
    5.1 After powering down, manually rotate the fan impeller to detect bearing seizure; replace the fan assembly if necessary.
    5.2 Inspect compressor shock-absorbing springs, pipe supports and screws; retighten any loosened parts with a torque wrench to specification.
    5.3 If metallic knocking is heard, stop the machine immediately to prevent catastrophic compressor failure.

    6 Other intractable problems
    If none of the above steps resolves the fault, record all alarm codes, pressures, currents, ambient conditions and the sequence of operations. Contact the manufacturer or an authorized service center at once for remote diagnosis or on-site repair. Never attempt unauthorized PCB soldering or pipeline modifications to avoid voiding the warranty.

    In short, troubleshooting a high-low temperature test chamber = proper operation + regular maintenance + systematic diagnosis + timely professional support. Follow these four principles and downtime will be minimized.

    News Recommendation
    If the high and low temperature humidity test chamber is used frequently and for a long time, the water in the equipment's water tank may not be sufficient to complete the test.
    The constant temperature and humidity test chamber is a common testing device used across many industries.
    When test samples are exposed to low temperatures in a thermal shock test chamber, they may suffer certain adverse effects. This is because the physical properties of the materials can change under extreme cold, potentially leading to performance degradation. So, what specific impacts can occur?
    A constant temperature and humidity test chamber is designed to maintain stable temperature and humidity conditions.
    A constant temperature and humidity test chamber is a device that can accurately simulate specific temperature and humidity environments. It is widely used in many fields such as electronics, automotive, aerospace, and biopharmaceuticals.
    Product Recommendation
    Telegram WhatsApp Facebook VK LinkedIn
    免费无码av片在线观看| 国产AV无码专区亚洲AV手机麻豆| 制服丝袜日韩中文字幕在线| 国产成人精品无码一区二区| 国产av无码专区亚洲国产精品| 无码毛片一区二区三区中文字幕| 日韩A无码AV一区二区三区| 国产精品一区二区久久精品无码| 久久AV高潮AV无码AV| 久热中文字幕无码视频| 99热门精品一区二区三区无码| 国产啪亚洲国产精品无码 | 十八禁视频在线观看免费无码无遮挡骂过 | 久久亚洲精品成人av无码网站| 玖玖资源站中文字幕在线| 亚洲无码精品浪潮| 国产真人无码作爱视频免费| 伊人久久无码中文字幕| 中文字幕成人精品久久不卡| 亚洲日本va中文字幕久久| 日韩乱码人妻无码中文字幕视频| 久久国产精品无码一区二区三区| 亚洲AV中文无码字幕色三| 高清无码v视频日本www| 色婷婷综合久久久久中文字幕| 亚洲中文字幕第一页在线 | 最好看更新中文字幕 | 熟妇人妻中文字幕无码老熟妇| 免费无码黄网站在线看| 潮喷失禁大喷水无码| 免费无码又爽又刺激一高潮| 日韩人妻无码一区二区三区| 高清无码午夜福利在线观看 | 成人午夜福利免费无码视频| 中文字幕亚洲一区| 最近中文字幕免费大全| 成人无码区免费A片视频WWW| 曰韩人妻无码一区二区三区综合部 | 92午夜少妇极品福利无码电影 | 亚洲综合无码一区二区| 国产乱妇无码大片在线观看|